• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 3FR SL PROVENA MIDLINE MAX BARRIER KIT; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 3FR SL PROVENA MIDLINE MAX BARRIER KIT; MIDLINE CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer "on (b)(6) 2022 one of the members was placing a 3 french bard provena power injectable midline on a patient.The nurse had no issues cannulating the vessel but when she attempted to dilate the vessel, she noticed that she was meeting resistance with the introducer/dilator.When she inspected the introducer/dilator, she noticed that the sheath of the introducer was pealing backwards.She obtained another introducer/sheath and was able to easily finish dilating the vessel and place the midline catheter.".
 
Additional Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of reft1128 showed no other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation:¿patient severity, applicable previous investigation(s), sample analysis, applicable fmea documents, applicable manufacturing records, and labeling.Based on a review of this information, the following was concluded: the complaint of a difficult insertion is confirmed.One 3.5 fr x 7cm microez microintroducer was returned for evaluation.The product sticker label indicated lot: reft1128.The dilator was returned, and the sheath had not been peeled.The sheath was observed to be bent at the proximal end extending from the peelable tabs and in the center region of the sheath tubing.Bunching of the sheath tip was also noted.Microscopic observation of the sheath tip revealed the material to be compressed and bunched inwards.The sheath tip taper transition did not appear to be smooth.The sheath tip surface was flat which created a blunt surface.Since the condition of the sheath tip taper likely contributed to the insertion of the device, the complaint is confirmed.Bd is working closely with the manufacturing facility to prevent reoccurrence of the reported event.
 
Event or Problem Description
It was reported by the customer "on (b)(6) 2022 one of the members was placing a 3 french bard provena power injectable midline on a patient.The nurse had no issues cannulating the vessel but when she attempted to dilate the vessel, she noticed that she was meeting resistance with the introducer/dilator.When she inspected the introducer/dilator, she noticed that the sheath of the introducer was pealing backwards.She obtained another introducer/sheath and was able to easily finish dilating the vessel and place the midline catheter.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3FR SL PROVENA MIDLINE MAX BARRIER KIT
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key14957665
Report Number3006260740-2022-02574
Device Sequence Number6915578
Product Code PND
UDI-Device Identifier00801741153952
UDI-Public(01)00801741153952
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K153393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberS4153108D
Device Lot NumberREFT1128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer10/06/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-