• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC Back to Search Results
Model Number MD17012601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
Bedside rn noticed leakage of blood tinged iv fluid around dressing.Picc team called and upon assessment noticed leaking at insertion site when flushed with ns.Removed picc and upon examination, noticed a ruptured area that was under the insertion site.
 
Additional Manufacturer Narrative
Our investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
The vascu picc was returned for evaluation.Visual inspection revealed two distinct kinks in the lumen, one at 7 cm from the hub and the other at 10 cm from the hub.Dried blood was removed from the distal end of the catheter.It cannot be determined if this was in place at the time of event or if it may have contributed to the event.Flushing the device through the luer with the red clamp no leaks were noted.However, when flushing through the luer with the white clamp a leak was noted at the 10 cm mark, where the device is kinked.Under magnification it is seen that the kink at 7 cm has rippled and bubbled, while the lumen has burst at the 10 cm kink.It can be seen that the burst was from the inside out.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.The device was in use for more than one month prior to the event with no reported issues.We are unable to determine the circumstances that may have led to this event.The instructions for use (ifu) contains the following warnings and precautions: do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).Only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.Bolus injections should be slow and must not exceed the maximum bolus pressure small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.Do not use sharp instruments near the extension lines or catheter lumen.Warning: do not flush against resistance.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL VASCU-PICC
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key14957693
Report Number2518902-2022-00043
Device Sequence Number12499010
Product Code LJS
UDI-Device Identifier00884908146493
UDI-Public884908146493
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K102966
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/01/2023
Device Model NumberMD17012601
Device Catalogue NumberMD17012601
Device Lot NumberMQDZ810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer06/22/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age7 YR
Patient SexFemale
-
-