Model Number 190858
Medical Device Problem Codes Thermal Decomposition of Device (1071); Sparking (2595)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t machine tripped the ground-fault circuit interrupter (gfci) power outlet and a spark could be seen from the power plug.Blown fuses were also identified on the power supply after the machine was pulled from service for evaluation after failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The manufacturer's investigation determined that there was no causal relationship between the objective evidence and the alleged event; therefore the alleged event is unconfirmed.
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t machine tripped the ground-fault circuit interrupter (gfci) power outlet and a spark could be seen from the power plug.Blown fuses were also identified on the power supply after the machine was pulled from service for evaluation after failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM WITHOUT CDX
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key14957759
Report Number0002937457-2022-01140
Device Sequence Number12136881
Product Code KDI
UDI-Device Identifier00840861100903
UDI-Public00840861100903
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K150708
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number190858
Device Catalogue Number190858
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-