| Model Number |
190858 |
| Medical Device Problem Codes |
Thermal Decomposition of Device (1071); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t machine tripped the ground-fault circuit interrupter (gfci) power outlet and a spark could be seen from the power plug.Blown fuses were also identified on the power supply after the machine was pulled from service for evaluation after failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The manufacturer's investigation determined that there was no causal relationship between the objective evidence and the alleged event; therefore the alleged event is unconfirmed.
|
| |
|
Event or Problem Description
|
|
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t machine tripped the ground-fault circuit interrupter (gfci) power outlet and a spark could be seen from the power plug.Blown fuses were also identified on the power supply after the machine was pulled from service for evaluation after failing to power on.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.Additional information was requested, however a response was not received.No parts were returned to the manufacturer for physical evaluation.
|
| |
|