U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER ACCUTNI+3 FOR USE ON ACCESS; IMMUNOASSAY METHOD, TROPONIN SUBUNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER ACCUTNI+3 FOR USE ON ACCESS; IMMUNOASSAY METHOD, TROPONIN SUBUNIT Back to Search Results
Model Number A98143
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/27/2021
Type of Reportable Event Death
Additional Manufacturer Narrative
The full identifier for this event is (b)(6).The access accutni+3 reagent and patient sample were not returned for evaluation.No hardware errors or issues with other assays were reported in conjunction with this event.No other patient results were called into question.System performance indicators such as system check, calibration and quality control were passing within specifications at the time of the event.The customer verbally confirmed that there is no indication that the accutni+3 result was erroneous.In conclusion, the cause of this event cannot be determined with the available information.
 
Event or Problem Description
On (b)(6) 2002 beckman coulter was made aware of a patient death potentially associated with the accutni+3 assay (access accutni+3, part number a98143 and lot number 125173) through notification from the fda.One patient sample was tested with the access accutni+3 assay; the sample generated a result which was within the assay's normal reference range.The result was not questioned by physician.Patient sample was not re-tested and no sequential samples from this patient were tested using the access accutni+3 assay following generation of the first accutni+3 result.The patient sample was not tested on any other methodologies and no serial draws were performed by the customer.There was a report of a death.The patient presented with reported chest pain and was treated in the emergency room on (b)(6) 2021, and was subsequently discharged.The patient passed away five days later, not at the hospital.The cause of death is unknown.No hardware errors or issues with other assays were reported in conjunction with this event.No other patient results were called into question.Per the hospital's verbal confirmation, system performance indicators such as system check, calibration and quality control were passing within specifications at the time of the event.Access accutni+3 reagent and calibrator lots met all specifications for release.It was indicated that the patient was under biotin supplementation, it is unknown which dose the patient was receiving.Per the accutni+3 instructions for use, part number b16315 j, the analytical specificity / interferences section, biotin was tested as a potential interfering substance and no interference was observed up to 290 ng/ml.The accutni+3 assay is not intended to be used in isolation; results should be interpreted in conjunction with other diagnostic tests and clinical information.No sample collection or processing information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUTNI+3 FOR USE ON ACCESS
Common Device Name
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
Manufacturer (Section G)
BECKMAN COULTER
1000 lake hazeltine drive
chaska MN 55318
Manufacturer Contact
angela vettel
250 s. kraemer blvd.
brea, CA 92821
7149613625
MDR Report Key14958054
Report Number2122870-2022-00031
Device Sequence Number5497633
Product Code MMI
UDI-Device Identifier15099590209322
UDI-Public(01)15099590209322(17)220930(11)210930(10)125173
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K121214
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberA98143
Device Catalogue NumberA98143
Device Lot Number125173
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Initial Report FDA Received Date07/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexMale
-
-