BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7214
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheter entrapment occurred.The target lesion was located in left anterior descending artery.A 3.75mm x 20mm nc emerge balloon catheter was advanced for dilatation.After balloon inflation, the physician attempted to remove the device.However, the device was stuck on the unknown guidewire.Both the balloon catheter and the wire were pulled as a unit.The procedure was completed with a different device.No patient complications were reported.
 
Event or Problem Description
It was reported that catheter entrapment occurred.The target lesion was located in left anterior descending artery.A 3.75mm x 20mm nc emerge balloon catheter was advanced for dilatation.After balloon inflation, the physician attempted to remove the device.However, the device was stuck on the unknown guidewire.Both the balloon catheter and the wire were pulled as a unit.The procedure was completed with a different device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The device was visually and microscopically examined.There was contrast present in the inflation lumen and the balloon.There was blood in the guidewire lumen and the balloon was loosely folded.The guidewire lumen was buckled and prolapsed 46mm from the tip at 5mm in length.Product analysis confirmed the reported event as the guidewire lumen was buckled and prolapsed.The damage is likely consistent with an interaction with a guidewire and could cause difficulty advancing and the wire to freeze up within the device during use or removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14958108
Report Number2134265-2022-07093
Device Sequence Number5395865
Product Code LOX
UDI-Device Identifier08714729846680
UDI-Public08714729846680
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141236
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2024
Device Model Number7214
Device Catalogue Number7214
Device Lot Number0028911662
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/27/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexMale
-
-