| Model Number |
1516-20-514 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Synovitis (2094); Joint Laxity (4526)
|
| Date of Event |
05/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for instability.Event is not serious and is considered moderate.Event is possibly related to both device and procedure.Date of implant: (b)(6) 2018.Date of revision: (b)(6) 2022.Date of event: (b)(6) 2020.Affected side: right knee.Treatment: revision, insert was removed.Depuy products: catalog id: 150400225; lot id: 8572102; component type: femoral; description: attune c/ret fem rt sz 5 narrow cemented.Catalog id: 150670004; lot id: 8648915; component type: tibial; description: attune fb cemented tibial base size 4.Catalog id: 151820038; lot id: 8641265; component type: patellar; description: attune medial dome pat 38mm.Catalog id: 151620514; lot id: j7644f; component type: insert; description: attune cr/fb insert sz 5 14mm.Competitor cement used.Insert was removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Event or Problem Description
|
|
Clinical database information indicated that "on (b)(6) 2022 the patient had a revision right total knee replacement to address pain and instability.The primary implants were well fixed with no signs of infection.The insert was revised.Depuy insert was used during this procedure.During the procedure the surgeon observed hemosiderin stained synovium.".
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|