Model Number 1550350-33
Medical Device Problem Codes Material Rupture (1546); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified, mildly tortuous lesion in the right coronary artery (rca).A 3.5x33mm xience sierra drug eluting stent (des) was advanced to the target lesion and inflated one time below nominal pressure when a rupture occurred.The stent was not fully apposed to the vessel wall and required additional dilatation to be fully implanted, therefore another 3.5mm balloon was used to post-dilate the stent and successfully complete the procedure.It was commented that the calcified part in the lesion or a stent that had already been deployed at the proximal part to the lesion, probably caused the rupture, but that it cannot be concluded.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported material rupture was confirmed.The balloon material was noted to be scratched, with a pinhole rupture.The reported difficult or delayed activation could not be tested due to the component not being returned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported difficulties; however, the subsequent treatment appears to be related to the operational context of the procedure.Factors that may contribute to material rupture include, but are not limited to, material damage, inflation technique, interactions with other devices, lesion calcification / tortuosity, and/or insufficient preparation prior to use.Factors that may contribute to difficult stent activation include, but are not limited to, patient¿s anatomical morphology, loose connection with the indeflator, and/or damage to the inflation lumen / balloon.In this case, it is possible that the stent delivery system (sds) interacted with a previously implanted stent and/or the moderately calcified and mildly tortuous anatomy during advancement, contributing to the noted scratched material (balloon), in addition to, the reported material rupture and subsequent difficult stent activation; however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14958186
Report Number2024168-2022-07374
Device Sequence Number6906213
Product Code NIQ
UDI-Device Identifier08717648227455
UDI-Public08717648227455
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/16/2023
Device Model Number1550350-33
Device Catalogue Number1550350-33
Device Lot Number0073041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-