Model Number 4540032-02
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun melsungen ag internal report (b)(4).The complaint is still under investigation.A follow-up report will be provided, as soon as investigation has been completed.
 
Event or Problem Description
As reported by the user facility / translation of user facility information by bbm sales organization in germany: "overinfusion" according to the customer: "easypump lt 2ml/h 270ml, ch.Ref.(b)(4).The easypump from the mentioned batch was supposed to run for 5 days (120h), filling 240ml with 5-fu/ nacl 0.9%.However, after about 3 days the pump was completely empty.No user errors are apparent, according to consultation the patient did not "lie on" the pump or otherwise exert pressure on the pump.".
 
Additional Manufacturer Narrative
This report has been identified as b.Braun melsungen ag internal report # (b)(4).As no sample and no meaningful picture was provided, proper investigation on malfunction could not be performed.Hence, the complaint is considered as not confirmed.Device history record (dhr): reviewed the dhr for batch 21g27ge341, there is no abnormality and no such defect detected at in process and at final control inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP® II
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key14958285
Report Number9610825-2022-00254
Device Sequence Number5831859
Product Code MEB
UDI-Device Identifier04046964448843
UDI-Public(01)04046964448843(17)260701(10)21G27GE341
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K081905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2026
Device Model Number4540032-02
Device Catalogue Number4540032
Device Lot Number21G27GE341
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer06/15/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-