| Model Number |
4540032-02 |
| Medical Device Problem Code |
Excess Flow or Over-Infusion (1311)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report (b)(4).The complaint is still under investigation.A follow-up report will be provided, as soon as investigation has been completed.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility / translation of user facility information by bbm sales organization in germany: "overinfusion" according to the customer: "easypump lt 2ml/h 270ml, ch.Ref.(b)(4).The easypump from the mentioned batch was supposed to run for 5 days (120h), filling 240ml with 5-fu/ nacl 0.9%.However, after about 3 days the pump was completely empty.No user errors are apparent, according to consultation the patient did not "lie on" the pump or otherwise exert pressure on the pump.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun melsungen ag internal report # (b)(4).As no sample and no meaningful picture was provided, proper investigation on malfunction could not be performed.Hence, the complaint is considered as not confirmed.Device history record (dhr): reviewed the dhr for batch 21g27ge341, there is no abnormality and no such defect detected at in process and at final control inspection.
|
| |
|