| Model Number |
3142 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is question was received and inspected.It was returned with 1/2 of heat shrink.Observed that heat shrink was shriveled indicating that the tip was reprocessed at the hospital.This most likely caused the heat shrink to crack/break in two length wise.
|
| |
|
Event or Problem Description
|
|
Name plate collar split in half intraoperatively unable to find one half of the piece.The dislodged (plastic portion) portion of the tip was never recovered after the procedure.Sales rep indicates that the tip split longitudinally.
|
| |
|