• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EZ GLIDE AORTIC CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EZ GLIDE AORTIC CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number EZC24TA
Medical Device Problem Code Nonstandard Device (1420)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
These devices are inserted into the blood vessels during cardiac surgery.Cracks in the material may lead to loose fragments in the vasculature which could embolize distally and cause infarction.The device was returned and product evaluation is ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event or Problem Description
It was reported that the vented connector of a ez glide aortic cannula came loose and the distal tip had a crack.There were no patient injuries.Per rep this was a cabg procedure and the event occurred towards the end of the procedure.No exchange of cannula was required.No abnormality of the device or packaging was noted prior to use.
 
Additional Manufacturer Narrative
Based on the information available, the complaint of cracked tip is confirmed, however, the most likely cause is due to excessive force applied by the user.An edwards defect has not been confirmed.
 
Event or Problem Description
It was reported that the tip of a ezc24ta ez glide aortic cannula was cracked.There were no patient injuries.Per rep this was a cabg procedure and the event occurred towards the end of the procedure.No exchange of cannula was required.No abnormality of the device or packaging was noted prior to use.Per the perfusionist the it came loose at the red-cap.It has to deal with the bonded vent connector site.The perfusionist feels the issue lies within the small vent connector barb and recommends a three barb connector.
 
Additional Manufacturer Narrative
Updated b5 per new information received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZ GLIDE AORTIC CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
one edwards way
mailstop anton 6.1
irvine, CA 92614
9492502731
MDR Report Key14959177
Report Number2015691-2022-06684
Device Sequence Number20182485
Product Code DWF
UDI-Device Identifier00690103714418
UDI-Public(01)00690103714418(17)241024(11)211025(10)328572
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K123370
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/24/2024
Device Model NumberEZC24TA
Device Catalogue NumberEZC24TA
Device Lot Number328572
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/01/2022
09/21/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date07/26/2022
09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient SexUnknown
-
-