| Model Number |
EZC24TA |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
These devices are inserted into the blood vessels during cardiac surgery.Cracks in the material may lead to loose fragments in the vasculature which could embolize distally and cause infarction.The device was returned and product evaluation is ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Event or Problem Description
|
|
It was reported that the vented connector of a ez glide aortic cannula came loose and the distal tip had a crack.There were no patient injuries.Per rep this was a cabg procedure and the event occurred towards the end of the procedure.No exchange of cannula was required.No abnormality of the device or packaging was noted prior to use.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information available, the complaint of cracked tip is confirmed, however, the most likely cause is due to excessive force applied by the user.An edwards defect has not been confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of a ezc24ta ez glide aortic cannula was cracked.There were no patient injuries.Per rep this was a cabg procedure and the event occurred towards the end of the procedure.No exchange of cannula was required.No abnormality of the device or packaging was noted prior to use.Per the perfusionist the it came loose at the red-cap.It has to deal with the bonded vent connector site.The perfusionist feels the issue lies within the small vent connector barb and recommends a three barb connector.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated b5 per new information received.
|
| |
|