• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED ONE STEP PEDIATRIC CPR ELECTRODE (1 PAIR); AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED ONE STEP PEDIATRIC CPR ELECTRODE (1 PAIR); AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 8900-000219-01
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the electrode pads would not adhere to the patient's skin.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.Please reference medwatch report numbers 1218058-2022-00065, 1218058-2022-00067, 1218058-2022-00059 for similar reports from the same customer.
 
Additional Manufacturer Narrative
The electrode pads were not returned to zoll for evaluation, however, three lot numbers, 1222, 0222e, and 0522c, were provided.The lot numbers are from the customer's inventory.It is not known if the electrode pads from the event were part of one of these lots.The customer did indicate one of the sites is currently using lot# 3912b.A dhr was performed on these lot numbers and no issue was found.However, without receipt of the electrode pads we are unable to conclusively understand the customer report or potential contributors.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONE STEP PEDIATRIC CPR ELECTRODE (1 PAIR)
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key14959390
Report Number1218058-2022-00066
Device Sequence Number5459330
Product Code MKJ
UDI-Device Identifier00847946023348
UDI-Public00847946023348
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number8900-000219-01
Device Catalogue Number8900-000219-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-