• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC THAL-QUICK ABSCESS DRAINAGE SET; GBX CATHETER, IRRIGATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THAL-QUICK ABSCESS DRAINAGE SET; GBX CATHETER, IRRIGATION Back to Search Results
Model Number N/A
Medical Device Problem Codes Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/21/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a thal-quick abscess drainage set had insufficient drainage.The device was required for a study procedure (mdr-1919) and was placed on (b)(6) 2021 for abscess drainage for a left side pneumothorax.Dilation was performed prior to catheter placement and computed tomography (ct) was utilized for visual guidance during the procedure.The drainage catheter was successfully placed and drainage was established.However, complete fluid evacuation was not achieved with this device.Six days post placement, on (b)(6) 2021, the device was removed due to insufficient drainage.The patient then required video assisted thoracoscopic surgery (vats) and pleurodesis to resolve the pneumothorax.No adverse were reported prior to the patient exiting the study.
 
Additional Manufacturer Narrative
Pma/510(k) #: exempt.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: it was reported by cri via baylor university medical center (usa) that a patient experienced insufficient drainage related to a thal-quick abscess drainage set (rpn: tqas-1200-j; lot#: 13977567).The device was required for a study procedure ((b)(6)) and was placed on (b)(6) 2021 for abscess drainage for a left side pneumothorax.Dilation was performed prior to catheter placement and computed tomography (ct) was utilized for visual guidance during the procedure.The drainage catheter was successfully placed, and drainage was established.However, complete fluid evacuation was not achieved with this device.Six days post placement, on (b)(6) 2021, the device was removed due to insufficient drainage.The patient then required video assisted thoracoscopic surgery (vats) and pleurodesis to resolve the pneumothorax to prevent permanent impairment or death.No additional adverse events were reported prior to the patient exiting the study.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures of the device, were conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document-based investigation evaluation was performed.Cook reviewed the device history record for lot 13977567 and records no non-conformances.The sub assembly lots (ni13977566 and sa13637720) were reviewed and found no non-conformances.A database search for complaints on the reported lot found no additional complaints reported from the field.The information provided upon review of dmr, and dhr, does not indicate the device was manufactured out of specification.There is no evidence of nonconforming material in house or in the field.Cook also reviewed product labeling.The product ifu, [t_tqas_rev9] ¿thal-quick abscess drainage sets,¿ provides the following information to the user related to the reported failure mode: precautions: ¿catheters should be irrigated on a routine basis to ensure function.Patient with indwelling drainage catheters should be evaluated routinely to ensure continuous function of the catheter.¿ instructions for use ¿13.Contrast may be injected under fluoroscopic visualization to verify positioning within abscess cavity.Note: it is important that all side ports of the drainage catheter are positioned within the abscess; the most proximal side port forms a break in the solid radiopaque stripe.In some large abscess cavities, it may be useful to insert two catheters, one small and one large, so that the saline may be irrigated through the small catheter and drainage collected through the large one.¿ based on the information provided, no returned device, and the results of the investigation, cook has concluded that component failure unrelated to manufacturing or design deficiencies contributed to the event.It is possible that the catheter wasn¿t irrigated routinely enough to ensure full function, but cook is unable to confirm this.It is also possible that patient anatomy and/or condition contributed to this event; however, these possibilities cannot be confirmed without additional information.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THAL-QUICK ABSCESS DRAINAGE SET
Common Device Name
GBX CATHETER, IRRIGATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14959639
Report Number1820334-2022-01179
Device Sequence Number17490248
Product Code GBX
UDI-Device Identifier00827002063365
UDI-Public(01)00827002063365(17)240818(10)13977567
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2024
Device Model NumberN/A
Device Catalogue NumberTQAS-1200-J
Device Lot Number13977567
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date01/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age51 YR
Patient SexMale
Patient Weight85 KG
-
-