• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION BACK-UP MEIER; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION BACK-UP MEIER; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 41488
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that guidewire coating peeled off.A.035in, 185cm, j tip, b5 back-up meier guidewire was used during the procedure.However, multiple places of the guidewire coating peeled off.The device was removed and no coating/wire fragments were left inside the patient's body.The procedure was completed successfully.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: unit guidewire was returned inside the original pouch of the device.The information of the label matches with the information of the complaint.Also, it was observed that the guidewire was inside the protective hoop.The pouch was wrinkled; and it shows that the hoop was manipulated.No damages were found in the returned device.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported that guidewire coating peeled off.A.035in, 185cm, j tip, b5 back-up meier guidewire was used during the procedure.However, multiple places of the guidewire coating peeled off.The device was removed and no coating/wire fragments were left inside the patient's body.The procedure was completed successfully.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BACK-UP MEIER
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key14959867
Report Number2134265-2022-07262
Device Sequence Number6945601
Product Code DQX
UDI-Device Identifier08714729227885
UDI-Public08714729227885
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K020283
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2024
Device Model Number41488
Device Catalogue Number41488
Device Lot Number0029315086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-