• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 5.4 MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® K2 EDTA (K2E) 5.4 MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367859
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tube there was foreign matter in the tube(s) biological and non-biological.The following information was provided by the initial reporter.The customer stated: there is "something long protruding inside tube wall.".
 
Additional Manufacturer Narrative
H6: investigation summary bd received 1 sample and 1 photo from the customer in support of this complaint.Evaluation of the sample and photo confirmed the presence of damage to the inside of the tube, on the side wall, has an indention and the plastic has been peeled away from the inside wall of the tube.No unused samples have been returned for evaluation.Nowhere on the production line is there anything introduced to the inside of the tube that would cause this type of defect.Additionally, 100 retention samples from the bd inventory were visually inspected with no damage to the inside of the tube being identified.The device history records were reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Bd was able to confirm the customer¿s indicated failure mode with the photos provided however, this defect is not caused by any process during the manufacturing of this product.The exact cause of the failure mode cannot be determined.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4 mg blood collection tube there was foreign matter in the tube(s) biological and non-biological.The following information was provided by the initial reporter.The customer stated: there is "something long protruding inside tube wall.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 5.4 MG BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14959879
Report Number1917413-2022-00374
Device Sequence Number5485912
Product Code JKA
UDI-Device Identifier50382903678599
UDI-Public50382903678599
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/28/2023
Device Model Number367859
Device Catalogue Number367859
Device Lot Number1288698
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2022
Supplement Date Received by Manufacturer07/18/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-