MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG EASYPUMP® II; ELASTOMERIC INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG EASYPUMP® II; ELASTOMERIC INFUSION PUMP Back to Search Results
Model Number 4540008-02
Medical Device Problem Codes Excess Flow or Over-Infusion (1311); Insufficient Flow or Under Infusion (2182)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility / translation of user facility information by bbm sales organization in australia: "overinfusion" reason of complaint: pump ran empty after 12 hours.Pump filled with ceftiaxone to infuse, over 24 hours.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun melsungen ag internal report (b)(4).The complaint is under evaluation.A follow-up report will be provided after the examination results are available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical, inc.Internal report (b)(4).The actual device involved in the reported incident was returned for evaluation.The reported issue of underinfusion was not present during investigation.Volumetrics of 100ml/hr and 10ml tested in spec at 9.9ml or 99% of expected volume.Perform preventive maintenance service.Log review confirms da 1222 error.
 
Additional Manufacturer Narrative
Event 1: this report has been identified as b.Braun medical internal report number (b)(4).No sample and/or lot number were provided.Further investigation of the complaint is not possible without a sample and/or lot number.The actual defective device is valuable tool in investigating the cause of this incident.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYPUMP® II
Common Device Name
ELASTOMERIC INFUSION PUMP
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsunge
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen 34212
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key14960951
Report Number9610825-2022-00270
Device Sequence Number20241838
Product Code MEB
UDI-Device Identifier04046964448485
UDI-Public(01)04046964448485(17)261001(10)21L11GE221
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Model Number4540008-02
Device Catalogue Number4540008
Device Lot Number21L11GE221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer06/29/2022
06/29/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date11/09/2022
11/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-