| Model Number |
ZA9003 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that while implanting the lens, the surgeon noticed the lens was torn near the trailing haptic.The lens was cut into three pieces and removed.A replacement lens same model and diopter size was inserted.There was no incision enlargement, no vitrectomy, and no sutures done.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 7/13/2022.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the complaint lens was received inside of the original lens case.Visual inspection under magnification revealed that the lens was received cut in half and had a detached haptic.The lens was cleaned and, tears were revealed on the optic body.Based on the return condition of the lens, no further product evaluation could be performed.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|