WRIGHTS LANE SYNTHES USA PRODUCTS LLC SLIDE/FIXED HAMMER 700 GRAMS; HAMMER, SURGICAL Back to Search Results
Model Number 03.010.056
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: patient identifier#: (b)(6).Reporter is a j&j sales representative.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2022 when striking the driving cap to insert the nail into the medullary canal, a piece of the handle on the hammer broke off.There was a patient involvement.No further information provided.This report is for one (1) slide/fixed hammer 700 grams.This is report 1 of 1 for complaint: (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the combined hammer 700g can be mounted f/35 was broken from the distal end of the phenolic handle.The remainder of the instrument shows significant dents, wear marks and impact marks.A dimensional inspection for the combined hammer 700g can be mounted f/35 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the combined hammer 700g can be mounted f/35 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.It is most likely that over 17 years of use and repeated thermal sterilization cycles has contributed to the handle breakage.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: review of the device history record(s) for part#: 03.010.056, synthes lot#: 4803819 showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIDE/FIXED HAMMER 700 GRAMS
Common Device Name
HAMMER, SURGICAL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES BRANDYWINE
1303 goshen parkway
west chester PA 19380
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key14961223
Report Number2939274-2022-02582
Device Sequence Number6944475
Product Code FZY
UDI-Device Identifier10886982067289
UDI-Public(01)10886982067289
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/03/2025
Device Model Number03.010.056
Device Catalogue Number03.010.056
Device Lot Number4803819
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/11/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date07/08/2022
Supplement Report FDA Received Date07/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN DRIVING CAP.
Patient SexUnknown
-
-