| Model Number |
2653 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the investigation facility to allow for an analysis to be performed.If the product is received for evaluation or new information is obtained, a follow up report will be submitted.Initial reporter phone # exceeded the maximum allowable characters, phone # is as follows: (b)(6).Device not returned.
|
| |
|
Event or Problem Description
|
|
The customer reported the device no longer "conducts" after a short time in operation.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.No physical damage was observed during visual inspection.The device failed continuity testing, broken wire at emitter solder joint assembly.The device failed functionality testing using a lab monitor and a "replace sensor" error message was displayed.Sensor led does not illuminate.Initial reporter phone # exceeded the maximum allowable characters, phone # is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported the device no longer "conducts" after a short time in operation.No patient impact or consequences were reported.
|
| |
|