| Catalog Number |
0117310 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was returned for evaluation.The foreign material noted by the customer is not identified to be an insect.Visual evaluation finds a small particulate that appears to be a residual fiber or cardboard material on the outside of the sterile package.The particulate is located above the seal of the sterile pouch.The sterile barrier of the product remains intact and the particulate is not in contact with the implant.Based on the sample evaluation and investigation performed, the root cause of the reported event is determined to be manufacturing process related.To date, this is the only reported complaint from this manufacturing lot of 711 units released for distribution in december, 2021.Sample evaluated.
|
| |
|
Event or Problem Description
|
|
As reported, when opening the 3dmax light mesh package during a hernia repair procedure on (b)(6) 2022, it was found that what appeared to be an insect was in the package.The device was not used in the case.As reported, there was no damage noted on the outer package/inner package.Another 3dmax light was used to complete the procedure.There was no patient injury.
|
| |
|