| Model Number |
TRM-375 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Seroma (2069); Lymphoma (3263)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.The reported events of "capsular contracture, baker grade iii" and "seroma-late" are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii and seroma-late.
|
| |
|
Event or Problem Description
|
|
Healthcare reported bilateral exchange from textured to smooth implants due to the patients concerns with the product.Healthcare professional additionally reported capsular contracture , baker grade ii/iii and seroma.This is for the left side.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "positive for anaplastic large cell lymphoma".Cd30+ biomarker was received.Alk- biomarker has not been received.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation:analysis of the returned device identified: weight to spec, deformation device and yellow particles in the shell.Air voids and cloudy in the gel observed after autoclave cycle.Base on the device analysis the final assessment is: no issues found related with the manufacturing process.Oncology contact information: (b)(6).
|
| |
|