Model Number 470183-14
Device Problem Break (1069)
Patient Problem Insufficient Information (4580)
Event Date 05/23/2022
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis (fa) replicated and confirmed the customer reported complaint that the permanent cautery hook had a broken tip.Fa found the primary of broken proximal clevis to be related to the customer reported complaint.The instrument was found to have the plastic proximal clevis broken.One of the ears of the clevis was broken off and was not returned with the instrument as a result of breakage.The size of missing piece is approximately 0.144¿ x 0.188¿ in length.No cable/wire damage or missing pins were observed.No thermal damage was found.The root cause of broken instrument proximal clevis is typically attributed to mishandling/misuse, such as excess force applied to the distal end of the instrument.A review of the procedure log showed the permanent cautery hook (part# 470183-14 / lot# n10190514 0116) was last used on (b)(6) 2022 on system sk1508.The instrument had 7 lives remaining.This complaint remains a reportable event due to the following conclusion: failure analysis confirmed the instrument's clevis was missing.This instrument is designed with a clevis pin on the distal end allowing articulation of the instrument assemblies they are holding together and is secured to the device by swaging.If the pin is not properly fused together, the pin could become dislodged and fall into the patient.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Event Description
It was reported that prior to the start of a da vinci-assisted surgical procedure, the permanent cautery hook had a broken tip.A backup instrument was used.The device was not used on the patient.Intuitive surgical, inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Type of Device
PERMANENT CAUTERY HOOK
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14973739
MDR Text Key303955615
Report Number2955842-2022-12885
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874112311
UDI-Public(01)00886874112311(10)N10190514
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470183-14
Device Catalogue Number470183
Device Lot NumberN10190514 0116
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2022
Date Manufacturer Received06/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-