Model Number 420207-10
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during da vinci-assisted hysterectomy surgical procedure, the jaw of the tenaculum forceps instrument was found to be broken.A backup instrument of a different type was used and the procedure was completed with no reported injury.Intuitive surgical, inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the tenaculum forceps instrument involved with this complaint and completed the device evaluation.Failure analysis investigations confirmed the customer reported complaint.Failure analysis found the primary failure of broken pitch cable to be related to the customer reported complaint.For clarification, the instrument was found to have a broken pitch cable at the distal end.The broken cable segment that contains the crimp was still installed in the clevis.Pitch cable breakage occurs when tensile load exceeds the ultimate strength of the material.The pitch cable construction is designed to optimize load and fatigue (cycling) characteristics.Variation in customer use conditions, procedure type, patient anatomy, product handling, instrument tip lengths, grip torque and manufacturing tolerances are a few variables which can influence pitch cable failure.A review of the instrument log for the tenaculum forceps instrument (pn# 420207-10 / lot# n10210118-223 associated with this event has been performed.Per logs, the instrument was last used on (b)(6) 2021 on system rsh0263 for approximately 11 minutes.The alleged event occurred on the 9th use of the instrument.No image or video clip for the reported event was submitted for review.This complaint is being reported based on the following conclusion: this instrument is designed with two pitch cables, each with a crimp at the distal end.If a pitch cable breaks at the distal end, a cable segment and/or the crimp could fall into the patient.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
TENACULUM FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14973740
Report Number2955842-2022-12886
Device Sequence Number5145823
Product Code NAY
UDI-Device Identifier00886874111659
UDI-Public(01)00886874111659(10)N10210118
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420207-10
Device Catalogue Number420207
Device Lot NumberN10210118 223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/06/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/09/2022
Initial Report FDA Received Date07/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexFemale
-
-