MAUDE Adverse Event Report: MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number H1-M
Device Problems Difficult to Advance (2920); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  malfunction  
Event Description
Physician was attempting to use a hawkone directional atherectomy along with a non-medtronic 6fr sheath, a 6mm spider fx guidewire and embolic protection during procedure to treat a little calcified plaque lesion in the left proximal popliteal artery (pop) with chronic total occlusion (cto-100%).The vessel was little tortuous.The vessel diameter and lesion length are 5mm and 100mm respectively.The vessel was pre dilated but not post dilated.Ifu was followed.Guidewire advancement/prolapse occurred.The device was advanced over bifurcation.The guidewire was hydrated at preparation.During withdrawal, minimal resistance was felt.The guidewire prolapse did not cause tip damage or embolization during procedure.The guidewire lumen was not torn and it did not lock-up on the catheter.It was reported that the hawkone would not pack.The device would turn on and shave but the packer would not slide forward into the nosecone.Device was withdrawn from patient but secondarily encountered wire prolapse.After the h1-m was discovered to not be packing, the physician attempted to withdraw the device but felt a small resistance at the sheath.Physician carried out fluoroscopy of the sheath and discovered a wire loop.The spider wire would not push forward to reduce the loop and the sheath began to accordion.The entire system was abandoned and manual pressure was held.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: the cutter was in the window as the device was being removed from the patient.This particular m device would not pack and so the blade would not advance into the nosecone with the thumbswitch.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: the wire had to be cut to free the device from the sheath.The lesion was treated with several passes of the h1-m device but, after it was discovered to not pack, the device was attempted to be removed.When wire wrap was terminally unsolvable, the sheath access was abandoned.The device was cleaned afterwards and it had picked up plaque in the nosecone, so the lesion was atherectomized but not followed up with any sort of balloon or dcb.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis the device was received with the thumbswitch positioned halfway between the ¿on¿ and ¿off¿ position the cutter was positioned approximately 8mm in the housing and is blocked by biologics present in the housing the thumbswitch was advanced but the cutter was stopped by the biologics.The housing was flushed as per ifu, but all the biologics could not be removed.The thumbswitch was advanced again but the cutter would only advance approximately 16mm in the housing before being was stopped by the remaining biologics image analysis the customer returned two images.Image 1 appears to show a device in a patient¿s vasculature.There appears to be evidence of a guidewire prolapse.Image 2 shows a hawkone device with a guidewire prolapse.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14974775
MDR Text Key303223092
Report Number9612164-2022-02579
Device Sequence Number1
Product Code MCW
UDI-Device Identifier00643169968332
UDI-Public00643169968332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH1-M
Device Catalogue NumberH1-M
Device Lot Number0010967815
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/05/2022
Initial Date FDA Received07/11/2022
Supplement Dates Manufacturer Received07/05/2022
07/11/2022
10/06/2022
Supplement Dates FDA Received07/11/2022
07/14/2022
10/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/03/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-