MAUDE Adverse Event Report: EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 4, 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 02-012-51-4009
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, approximately 4 years port op the initial right tka, this 52 y/o male patient was revised due to poly wear and the poly being slightly delaminated.Parts and pieces were removed from patient by the surgeon, he pulled the poly out and irrigated the wound w 3000 cc of antibiotic.There was a 5-15 min surgical delay, no adverse event as a result.Patient was last known to be in stable condition following the event.Device is not returning.
 
Additional Manufacturer Narrative
(h3) the revision reported may have been due to a combination of orientation of the components, third body wear, and/or patient related conditions which led to wear of the tibial insert.However, this cannot be confirmed because the component was not returned for evaluation and images were not provided.The following sections have corrected information: (d1) brand name: logic tib insert impl crc, sz 4, 9mm.(d4) catalog number: 02-012-51-4009, serial number: (b)(6), expiration date: 28-dec-2023, unique identifier (udi)#: (b)(4).(h4) device manufacture date: 30-dec-2015.(h6) component code: 4755, part/component/sub-assembly term not applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC TIB INSERT IMPL CRC, SZ 4, 9MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key14975712
Report Number1038671-2022-00776
Device Sequence Number5099386
Product Code JWH
UDI-Device Identifier10885862230485
UDI-Public10885862230485
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K123342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/28/2023
Device Model Number02-012-51-4009
Device Catalogue Number02-012-51-4009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/30/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexMale
-
-