• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION®; DIMENSION® EXL¿ WITH LM INTEGRATED CHEMISTRY SYSTEM AND SAMPLE TRANSFER MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION®; DIMENSION® EXL¿ WITH LM INTEGRATED CHEMISTRY SYSTEM AND SAMPLE TRANSFER MODULE Back to Search Results
Model Number DIMENSION® EXL¿ WITH LM INTEGRATED CHEMISTRY SYSTEM AND SAMPLE TRANSFER MODULE
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
Five (5) discordant falsely depressed sodium (na) patient sample results were obtained on a dimension exl w/lm stm system.The discordant results were reported to the physician(s).The same samples were reprocessed on the same dimension system and higher results, considered correct, were obtained.Corrected reports were issued.There are no known reports of patient intervention or adverse health consequences due to the falsely depressed sodium results.
 
Additional Manufacturer Narrative
A united states customer contacted the siemens healthcare diagnostics remote services center (rsc) regarding five (5) falsely depressed sodium (na) patient sample results obtained on a dimension® exl¿ with lm integrated chemistry system and sample transfer module.Siemens is investigating the event.
 
Additional Manufacturer Narrative
Additional information (12-jul-2022): siemens healthcare diagnostics headquarters support center (hsc) concluded the investigation of the event.Hsc evaluated the information provided by the customer and the instrument data logs.The customer performed maintenance that was part of normal troubleshooting.The customer adjusted the dilution factor and processed quality control which was within laboratory acceptance range after maintenance.Repeat of the same samples yielded expected results.A potential product problem was not identified with the sodium assay or the dimension exl with lm system and sample transfer module.The cause of the event is unknown.The instrument is operational.The device is performing within specifications.No further evaluation is required.Section h6 has been updated to reflect the hsc investigation.Initial mdr 2517506-2022-00178 was filed 11-jul-2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION®
Common Device Name
DIMENSION® EXL¿ WITH LM INTEGRATED CHEMISTRY SYSTEM AND SAMPLE TRANSFER MODULE
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC. BROOKFIELD
registration number 1226181
101 silvermine road
brookfield CT 06804
Manufacturer Contact
james morgera
500 gbc drive
po box 6101
newark, DE 19714-6101
MDR Report Key14976451
Report Number2517506-2022-00178
Device Sequence Number5689853
Product Code JJE
UDI-Device Identifier00630414593593
UDI-Public00630414593593
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K130276
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDIMENSION® EXL¿ WITH LM INTEGRATED CHEMISTRY SYSTEM AND SAMPLE TRANSFER MODULE
Device Catalogue NumberSMN 10639337
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/16/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-