|
According to the reporter, following a procedure to transfer pulmonary vein to left atrium by primary suture method and patching of atrial septal defect, the patient was hemodynamically stable with open sternum.One hour following the surgery, the patient had bled profusely through the drainage and blood pressured dropped to approximately 10mmhg.The patient was given fluid and acute blood transfusion, intensive cardiopulmonary resuscitation and the patient was reopened at the chest.The primary suture line was seen to have broken in the middle.The surgeon tried to suture the atrial suture with size 7.0 other polypropylene suture, but could not suture.The patient died from acute blood loss.
|
|
D10.Concomitant prod: vp-630-x (brand name: vp-630x surgipro*ii 7-0 blu 60cm (b)(6), lot number: d1g0784y); vp-630-x (brand name: vp-630x surgipro*ii 7-0 blu 60cm (b)(6), lot number: d1g0784y); vp-630-x (brand name: vp-630x surgipro*ii 7-0 blu 60cm (b)(6), lot number: d1g0784y); vp-630-x (brand name: vp-630x surgipro*ii 7-0 blu 60cm (b)(6), lot number: d1g0784y); vp-630-x (brand name: vp-630x surgipro*ii 7-0 blu 60cm (b)(6), lot number: d1g0784y); h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device used in primary procedure was not returned.A representative device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the suture thread broke.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the represented device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|