• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. EXCALIBUR, CURVED 4.0MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. EXCALIBUR, CURVED 4.0MM X 13CM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Model Number EXCALIBUR, CURVED 4.0MM X 13CM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6)2022, it was reported by a sales representative via email that an ar-8400cex excalibur broke while cleaning up the tibial stump.The excalibur broke while inside the sheath and all the broken fragments were accounted for.A new ar-8400cex excalibur was opened and case was completed without further issues.This was discovered during an aclr procedure on (b)(6)2022.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint is confirmed.One unpackaged ar-8400cex was received for evaluation.Visual inspection identified the flexible coil connecting the cutting tip to the inner shaft was twisted and broken.A fragment of hard bone was found stuck inside the device.Without additional information, the cause remains undetermined.A probable cause is attributed to the hard bone fragment becoming lodged in the rotating cutting tip, and the torque from the shaft causing the flexible coil to break, rendering the device inoperable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCALIBUR, CURVED 4.0MM X 13CM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key14976660
Report Number1220246-2022-05204
Device Sequence Number5035653
Product Code GFA
UDI-Device Identifier00888867043503
UDI-Public00888867043503
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEXCALIBUR, CURVED 4.0MM X 13CM
Device Catalogue NumberAR-8400CEX
Device Lot Number14919235
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer06/23/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-