CARDINAL HEALTH 12FR 43 IRIS FEEDING TUBE ENF; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 461243E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that they were called to replace dht due to leaking.They arrived to find the tube feed line attached to the dht port.That port was cracked away from the main tube with only a small section of plastic holding it together.The tube feeding line was stuck in the hub and they were unable to disconnect it from the dhr.The old dht and bridle was removed and a new dht was placed.
 
Additional Manufacturer Narrative
A device history record review could not be performed because the serial and lot numbers were not received with the complaint.However, as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to the release of product.There were no photos or physical sample returned for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported condition and determine the root cause.If a sample is received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.A corrective action is not applicable at this time.Functional testing and visual inspections are being performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12FR 43 IRIS FEEDING TUBE ENF
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14977271
Report Number1282497-2022-10704
Device Sequence Number5098362
Product Code KNT
UDI-Device Identifier20884521742229
UDI-Public20884521742229
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number461243E
Device Catalogue Number461243E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/08/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient Age79 YR
Patient SexUnknown
-
-