This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.However, a photo was provided.Upon visual inspection of the photo, it appeared that the inner came off the shaft, the photo do not provide enough evidence to confirm the issue reported.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed.Without a physical evaluation of the device, we cannot determine the root cause.The possible root cause of the condition of the device could be related to heavily use; moreover, drop of the unit, mishandling or procedural variables that may cause damage to the device.Since this is a reusable device, this condition has possibly occurred after using it for many procedures.However, it cannot be conclusively affirmed.As per ifu, end of useful instrument life is generally determined by wear or damage from handling or surgical use.Inspect instruments between uses to verify proper functioning.Visually inspect the instrument and check for damage and wear, cutting edges should be free of nicks and have a continuous edge, jaws and teeth should align properly, moveable parts should have smooth movement without excessive play, locking mechanisms should fasten securely and close easily, long, thin instruments should be free of bending and distortion in accordance with the manufacturer¿s instructions.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
|