MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) IDENTITY ADX VDR; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) IDENTITY ADX VDR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5480
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented in clinic for a follow-up.Upon review, it was discovered that the pacemaker exhibited premature battery depletion.The device was explanted and replaced.Further information was requested however, was not provided.
 
Additional Manufacturer Narrative
The reported event of premature discharge of battery was not confirmed.The device was received from the field with battery voltage at elective replacement level triggering backup operation and reaching end of service level during the analysis.Review of the device image indicates there was a false replacement alert in the field consistent with low voltage fluctuation.Electrical analysis performed under nominal conditions, after replacing the battery, indicated normal functionality.Further evaluation at various pulse amplitudes measured current level within normal range, and no indication of premature discharge of battery noted.Longevity assessment was performed, based on field settings, and the device exceeded projected longevity indicating normal battery depletion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITY ADX VDR
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14979230
Report Number2017865-2022-14525
Device Sequence Number5073629
Product Code LWP
UDI-Device Identifier05414734006927
UDI-Public05414734006927
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P880086
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model Number5480
Device Catalogue Number5480
Device Lot NumberS000011804
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer08/26/2022
09/06/2022
09/29/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/26/2022
09/06/2022
09/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-