| Catalog Number |
03333701190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with alp2 alkaline phosphatase acc.To ifcc gen.2 on a cobas 6000 c (501) module (serial number (b)(4)).No incorrect results were reported outside of the laboratory.The sample initially resulted in an alp value of 386 u/l and repeated as 316 u/l.The sample was re-centrifuged twice and repeated five times with all results around 200 u/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation determined the issue is consistent with a high leukocyte concentration in the sample, leading to an accumulation of cells close to the plasma surface.
|
| |
|