ROCHE DIAGNOSTICS ALP2 ALP IFCC GEN.2; NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES Back to Search Results
Catalog Number 03333701190
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they received discrepant results for one patient sample tested with alp2 alkaline phosphatase acc.To ifcc gen.2 on a cobas 6000 c (501) module (serial number (b)(4)).No incorrect results were reported outside of the laboratory.The sample initially resulted in an alp value of 386 u/l and repeated as 316 u/l.The sample was re-centrifuged twice and repeated five times with all results around 200 u/l.
 
Additional Manufacturer Narrative
The investigation determined the issue is consistent with a high leukocyte concentration in the sample, leading to an accumulation of cells close to the plasma surface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALP2 ALP IFCC GEN.2
Common Device Name
NITROPHENYLPHOSPHATE, ALKALINE PHOSPHATASE OR ISOENZYMES
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key14979694
Report Number1823260-2022-02034
Device Sequence Number5098431
Product Code CJE
UDI-Device Identifier04015630919505
UDI-Public04015630919505
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K033185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number03333701190
Device Lot Number63164401
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Supplement Date Received by Manufacturer07/12/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-