Model Number MCH-1000(M)
Device Problems Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/02/2019
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.Cardioquip inspected the device and found the burning smell was caused by an overheated pump motor.Further inspection showed the pump motor overheating was caused by a cooling fan malfunction that was, in turn, caused by dust and debris.The fans were cleaned, pump motor lubricated and upon final inspection the device was fully functional.
 
Event Description
Customer reports that the unit has a burning smell while running with water, and that the motor gets very hot.
 
Manufacturer Narrative
Further review determined that this is a non-reportable event.
 
Event Description
Customer reports that the unit has a burning smell while running with water, and that the motor gets very hot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key14981920
MDR Text Key304551238
Report Number3007899424-2022-00052
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000(M)
Was Device Available for Evaluation? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-