BECTON DICKINSON MEDICAL (SINGAPORE) BD INSYTE-W¿ PERIPHERAL VENOUS CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381323
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that 1 bd insyte-w¿ peripheral venous catheter had needle through the catheter issues.The following information was provided by the initial reporter translated from french to english : the user reported that during a catheterization of a dog with a catheter insyte blue thread 0.9x25mm, when starting to remove the mandrel without taking it out completely there was no blood return so it was put back in and realized that it was blocked and a potato had formed at the vein.On removing the catheter it was notice that the mandrel is blunt and had pierced the catheter.
 
Event or Problem Description
It was reported that 1 bd insyte-w¿ peripheral venous catheter had needle through the catheter issues.The following information was provided by the initial reporter translated from french to english : the user reported that during a catheterization of a dog with a catheter insyte blue thread 0.9x25mm, when starting to remove the mandrel without taking it out completely there was no blood return so it was put back in and realized that it was blocked and a potato had formed at the vein.On removing the catheter it was notice that the mandrel is blunt and had pierced the catheter.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 05-jul-2022.H6: investigation summary: one actual sample with open packaging were received by our quality team for evaluation.The sample was subjected to visual inspection to check for needle pierced through the catheter.From the returned sample, the needle was observed to be pierced through the catheter near the tip area.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The assembly process was reviewed.If the needle pierced through catheter occurred in the manufacturing process, the defect would be detected and auto rejected by the inline tip spear vision inspection system.Needle pierced through catheter could also happen during product application during the needle cover removal if not done carefully.As the actual sample was returned in open packaging, the root cause could not be established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE-W¿ PERIPHERAL VENOUS CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14982770
Report Number8041187-2022-00356
Device Sequence Number5099016
Product Code FOZ
UDI-Device Identifier00382903813230
UDI-Public00382903813230
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151698
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date08/31/2026
Device Catalogue Number381323
Device Lot Number1230787
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-