It was reported that the procedure was to treat a mildly tortuous, mildly calcified internal carotid artery that was 80% stenosed.The emboshield nav6 was prepped accordingly without issue (filter loaded into the delivery catheter (dc) pod).The dc was advanced to the target lesion, and when deployment was initiated, it was noted that the tip of the dc was separated into two parts (still remaining on the barewire) and the filter completely failed to be deployed.The nav6 was simply withdrawn (including separated portion) and replaced with another new nav6 to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
Visual analysis was performed on the returned product.The reported separation of the dc pod which likely contributed to deployment failure was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.The investigation was unable to determine a cause for the reported difficulties.Based on the reported information and evaluation of the returned unit, the deployment failure appears to be the result of the separated distal shaft/dc pod.Although the cause for the separation could not be determined, it may be possible that the distal shaft of the delivery catheter was restricted or entrapped in the anatomy and during positioning, the separated occurred resulting in deployment failure; however, this could not be confirmed.There was no difficulty noted during preparation, and the filter was noted to be fully loaded into the dc pod suggesting that the damage was not pre-existing.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|