• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION ANGIOSCULPT RX PTCA; CATHETER, PTCA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION ANGIOSCULPT RX PTCA; CATHETER, PTCA, CUTTING/SCORING Back to Search Results
Model Number 2200-2010
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
The angiosculpt device was used in a severely calcified mid lad.During inflation at 17 atm, the balloon ruptured.The procedure was completed with a non-angiosculpt device.No patient injury reported.This product problem is being submitted conservatively because the balloon ruptured below rbp.
 
Additional Manufacturer Narrative
The patient''s dob, weight, ethnicity, and race are unknown.This information was not available from the facility.Report source: foreign- taiwan.The angiosculpt device was infectious and discarded by the facility, thus no returned product investigation was performed.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT RX PTCA
Common Device Name
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer Contact
diana melliza galvez
5055 brandin court
fremont, CA 94538
MDR Report Key14983068
Report Number3005462046-2022-00027
Device Sequence Number5728484
Product Code NWX
UDI-Device Identifier00813132021146
UDI-Public00813132021146
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
P050018
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/14/2023
Device Model Number2200-2010
Device Catalogue Number2200-2010
Device Lot NumberG20050014
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Initial Report FDA Received Date07/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASAHI: 0.014" SION BLUE GUIDEWIRE; TERUMO: 6F GUIDE CATHETER
Patient Age68 YR
Patient SexMale
-
-