| Model Number |
120-500 |
| Medical Device Problem Codes |
Material Rupture (1546); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Swollen Lymph Nodes/Glands (4432)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of lymphadenopathy is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: pain and lymphadenopathy.
|
| |
|
Event or Problem Description
|
|
Patient reported right side pain in armpits on right-side and lymphadenopathy.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported "right side rupture, soft tissue excess, waterfall deformity and exchange.".
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported "textured".The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: pain: unable to confirm as it is a medical event not related to the device.Lymphadenopathy: unable to confirm as it is a medical event not related to the device.Rupture: observed broken sharp assessed as an unidentified (tear) opening.Other-medical: unable to confirm as it is a medical event not related to the device.Visibility/palpability: unable to confirm as it is a medical event not related to the device.Additional observations: crease flat observed in the device.Wear abrasion observed in the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "pain in armpits on right-side" and "lymphadempathy".Healthcare professional additionally reported "right side rupture, soft tissue excess, waterfall deformity and exchange." healthcare professional additionally reported "textured".The device has been explanted and replaced.
|
| |
|