• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH ENTRFLX 10FR 43IN W STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENTRFLX 10FR 43IN W STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721088
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the guided wire nasal enteral probe did not come out in the passage of it.They had to remove it from the patient and pass another probe.The green protective cap fell and almost pinched the customer.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported issue.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.The device was manufactured on august 17, 2020.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform functional and visual evaluations to confirm the reported condition and determine the root cause.If a sample is received at a later date, the complaint will be reopened, and the investigation will be updated accordingly.A corrective action is not applicable at this time.Functional testing and visual inspections are performed according to our current quality standards and inspection procedures.This complaint will be used for tracking and trending purposes.Corrected section d9.Device available for evaluation? to reflect no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRFLX 10FR 43IN W STY YPORT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key14983281
Report Number9612030-2022-03309
Device Sequence Number14695829
Product Code KNT
UDI-Device Identifier10884521518872
UDI-Public10884521518872
Combination Product (Y/N)N
Initial Reporter CountryBR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884721088
Device Catalogue Number8884721088
Device Lot Number2022315464
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer06/16/2022
Initial Report FDA Received Date07/11/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-