As reported: pronto v4 5.5 fr used in an acute stemi for thrombectomy.Patient died before successful treatment of the infarction.Patient presented with acute mi.Transcatheter thrombectomy attempted with pronto v4 5.5 fr.While attempting to withdraw pronto catheter the catheter was reported to be stuck in a previously placed coronary stent.Chest compressions were performed as patient began to decline.It is not clear whether this was in relation to the original cause of the mi or due to complications of having a catheter stuck in the coronary.Later the pronto was removed and it was found that the coronary stent had been dislodged.Additional on 20jun2022 - the account reported that the pronto catheter was stuck in a stent that was placed approx.5 years ago.Additional information on 21jun2021: plavix was stopped before due to hip surgery.Patient had presented with a stemi on 16th into 17th (11pm - 1am).Thrombus was in circumflex mid distal artery.The previously placed stent size was 3.0mm and had in-stent restenosis.The pronto v4 catheter wouldn't go forward or back once they advanced the catheter the 3rd time after two successful passes.Nitro was given.A second guidewire was placed down, and ballooning was done to try to dislodge.During the process of cpr, everything came out of the body, including the guide catheter, pronto and the sheath.Pericardial tamponade was reported to be the result leading to patient's death.
|
|
One unit of pronto aspiration catheter along with guide catheter, hemostasis valve, guidewire and extension were returned.Blood particulates were noted on the all the units.The units were decontaminated and inspected.The stent strut was observed to embedded within the silva tip.The stent was deformed completely in its length.The stent was dismantled from the tip exposing the polymer of the silva tip being slit/damaged.All other components were dismantled from the pronto catheter except the guidewire.The guidewire was locked within the lumen.No other damages were observed.Note: while attempting to remove the guidewire from the pronto shaft during product evaluation, resistance was observed.This caused tip to be separated.The top-level assembly traveler was reviewed.There are no dda's, nce's or any other nonconformances related to this lot therefore supporting the device met material, assembly, and performance specifications.Case details were reviewed.An 81y/o female patient underwent a thrombectomy procedure for acute stemi.Pronto v4 5.5f aspiration catheter was used in the case.Account stated that while withdrawing the pronto catheter, it got stuck in a previously placed stent.The strut was dismantled from the tip that exposed the polymer of the silva tip to be partially slit and damaged.The stent was deformed completely in its length.The damages are likely to have occurred during use of the pronto when operated against resistance with the stent strut.Per ifu states the following warning and precaution: - never advance or withdraw an intravascular device against resistance until the cause of the resistance is determined by fluoroscopy.Movement of the catheter or guidewire against resistance may result in separation of the catheter or guidewire tip, damage to the catheter, or vessel perforation.- exercise care in handling of the catheter during a procedure to reduce the possibility of accidental breakage, bending or kinking.Additional information was requested from the account.A response was received.Thrombus was noted in the circumflex mid distal artery.Pronto v4 catheter wouldn't go forward or back once they advanced the aspiration catheter the 3rd time after two successful passes.A second guidewire was placed down, and ballooning was done to try to dislodge.The 3.0mm stent in which the pronto was stuck was placed 5 years ago.Per event details, chest compressions (cpr) were performed as patient began to decline.It is not clear whether this was in relation to the original cause of the myocardial infarction (mi) or due to complications of having a catheter stuck in the coronary.Per ifu vessel dissection, perforation, rupture, and mi are identified as potential adverse effects that maybe associated with pronto v4 aspiration catheter.Coronary stent was dislodged and removed along with pronto.Four fluoroscopic images were provided by the account without any description.The following were noted: a percutaneous ventricular assist device (pvad) was observed.It is unknown when the pvad was introduced into the procedure.Balloon inflation can be observed.Two guidewires were observed to be within left circumflex branches which likely represents ballooning attempt to dislodge tip.The image indicates possibility of dissections of the vessel though no confirmation from the account was provided.It is unclear as to where the pronto tip was present.The stent strut marks are not distinctly visible to confirm the issue.Procedural outcome was pericardial tamponade leading to patient's death.Based on the information, the most likely root cause of the issue is user error.
|