It was reported that during an arthroscopy, the biorci screw tip got broken while screwing in, all broken pieces were removed from the bone with the help of a grasper.The procedure was completed without significant delay using a s+n back-up device.No further complications were reported.
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.There was no relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.This case reports, the tip of the screw broke during implantation while screwing it in.Reportedly, all of the broken pieces were removed from the bone using a grasper.Smith and nephew has not received the device and/or adequate materials to fully evaluate the complaint.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.The complaint was not confirmed.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|