Information received via medtronic indicated that the customer experienced hyperglycemia.Customer's blood glucose value at time of incident was 312 mg/dl.Customer did experience symptoms related to high blood glucose such as nausea, vomiting, abdominal pain and difficulty breathing.Customer was treated with manual injection.Customer's current blood glucose value was 312 mg/dl.Customer was neither in emergency room, nor admitted into hospital, nor was emergency medical service dispatched.Customer did not allege that the insulin pump was under delivering.The device will be returned for analysis.
|
|
This report is part of a retrospective review and remediation efforts in response to a warning letter.Recall: z-0955-2020.(b)(4).S/w 4.10c.Unit passed the displacement test, rewind, prime/seating test, basic occlusion test and force sensor test.Unit received with high sleep current measurement due to corroded battery.Unit received with cracked retainer and pillowing keypad overlay.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ""defects"" or has ""malfunctioned"".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.".
|