ABBOTT MEDICAL AMPLATZER DUCT OCCLUDER II; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number DUCT OCCLUDER II
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that an amplatzer duct occluder ii 5mm dia 6mm was selected for an implant on (b)(6) 2022.During the procedure, device disc was not opening up properly.Device was removed from the patient prior to release from the delivery cable.There is no interaction with cardiac structures during deployment and no angulation or kink noticed in the delivery system.A non-abbott device was selected and implanted.The patient was reported to be in stable condition.
 
Additional Manufacturer Narrative
An event of device deformity was reported.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.Some possible causes of device deformity include use of an incorrect size sheath, anatomical interference, and angulation/bends in the sheath.Information from the field indicated that there was no interaction with cardiac structures or any angulation or kink in the delivery system.The field also indicated that the correct size delivery system was used.The cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There were no other complaints on this batch.There is no indication of a product quality with respect to labeling, design, or manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER II
Common Device Name
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key14983765
Report Number2135147-2022-00483
Device Sequence Number12220393
Product Code MAE
UDI-Device Identifier00811806011240
UDI-Public00811806011240
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P020024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2022
Device Model NumberDUCT OCCLUDER II
Device Catalogue Number9-PDA2-05-06
Device Lot Number6072911
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/13/2022
Supplement Date Received by Manufacturer10/12/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-