• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER; CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY Back to Search Results
Model Number D128210
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Cardiac Tamponade (2226)
Date of Event 06/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old male patient underwent an paroxysmal atrial fibrillation (afib) ablation procedure with a pentaray nav high-density mapping eco catheter.The patient suffered cardiac tamponade requiring pericardiocentesis.It was reported that prior to ablation, during mapping in the left atrium with the pentaray, anesthesia personnel noted hypotension.Using fluoro imaging, they noted that the heart border was not moving as normal.Ice (intracardiac echocardiography) was used to check the heart, and a pericardial effusion was noted.Pericardiocentesis was performed, and fluid amount withdrawn is unknown as it was auto-transfused back to the patient.This adverse event was discovered during use of bwi products.The physician¿s opinion on the cause of this adverse event is unknown.The patient outcome of the adverse event is stable.The patient did require extended hospitalization because of the adverse event.Two transseptal punctures were performed with a cook needle.Ablation was not performed prior to noting the pericardial effusion.There was no evidence of steam pop as no ablation was performed.The event occurred during mapping.No error messages were observed on biosense webster equipment during the procedure.The patient is hemodynamically stable at this time.The effusion was discovered with ice imaging after hypotension.The patient was admitted.Suspected perforation is unknown.It was confirmed with soundstar.Pericardiocentesis was performed.The tube was left in place.The procedure was abandoned.There was no evidence of any effusion present before the procedure.No ablation was performed so ablation catheter not opened.The physician¿s opinion on the cause of this adverse event reason is unknown.It could be due to anatomical structure.A pericardiocentesis was performed and the patient outcome of the adverse event was fully recovered.Since the event is life threatening and required intervention and prolonged hospitalization to prevent permanent impairment of a body function or permanent damage to a body structure, is to be considered serious and mdr-reportable.
 
Additional Manufacturer Narrative
On 12-aug-2022, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that a 76-year-old male patient underwent an paroxysmal atrial fibrillation (afib) ablation procedure with a pentaray nav high-density mapping eco catheter.The patient suffered cardiac tamponade requiring pericardiocentesis.Device evaluation details: visual analysis of the returned sample revealed that no damage or anomalies were observed on the pentaray nav eco catheter.Per the event, several tests were performed.The magnetic features were tested, and no issues were observed.In addition, the product was deflecting and irrigating correctly.No malfunctions were observed during the product analysis.A manufacturing record evaluation was performed for the finished device number 30713668l, and no internal actions related to the complaint were found during the review.No malfunction was observed during the product analysis.The instructions for use (ifu) states that careful catheter manipulation must be performed to avoid cardiac damage, perforation, or tamponade.The root cause of the adverse event remains unknown.There may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the analysis.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTARAY NAV HIGH-DENSITY MAPPING ECO CATHETER
Common Device Name
CATHETER, INTRACARDIAC MAPPING, HIGH-DENSITY ARRAY
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key14984015
Report Number2029046-2022-01577
Device Sequence Number12048466
Product Code MTD
UDI-Device Identifier10846835012248
UDI-Public10846835012248
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/06/2025
Device Model NumberD128210
Device Catalogue NumberD128210
Device Lot Number30713668L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/12/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date09/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SOUNDSTAR CATHETER
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age76 YR
Patient SexMale
-
-