| Catalog Number |
06368697190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter complained of discrepant high results for 1 patient sample tested for elecsys anti-tg (anti-tg) on two cobas 6000 e 601 modules compared to an unspecified method from another hospital.On (b)(6) 2022 the result from the e601 module was 510.600 iu/ml.This result was reported outside of the laboratory.On (b)(6) 2022 the patient went to another hospital to check the result.The result at the other hospital was 60.02 iu/ml.On (b)(6) 2022 the customer sent the original sample to another hospital and the result was 58.50 iu/ml.On (b)(6) 2022 the sample was repeated on the e601 module in question and the result was 583.6 iu/ml.The customer ran the sample on another e601 module and the result was 552.1 iu/ml.The serial number for one e601 module was 26e4-23.The serial number for the other e601 module was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Clarification was received that the anti-tg results originally reported of 60.02 iu/ml and 58.50 iu/ml were from a different sample from a different patient.The anti-tg results of 510.600 iu/ml, 583.6 iu/ml and 552.1 iu/ml were the results from the sample in question for the patient in this case.These results are not discrepant.The investigation did not identify a product problem.
|
| |
|