| Model Number |
PHOS2 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Na.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with phos2 phosphate (inorganic) ver.2 on a cobas c 503 analytical unit (serial number (b)(4)).No incorrect results were reported outside of the laboratory.The sample initially resulted in a phosphorus value of 2.69 mg/dl accompanied by a data flag indicating atypical kinetic behavior.The sample was repeated, resulting in a phosphorus value of 2.68 mg/dl accompanied by a data flag indicating atypical kinetic behavior.The sample was diluted 1:3 and repeated, resulting in a phosphorus value of 2.33 mg/dl accompanied by a data flag.The sample was repeated again at decreased sample volume, resulting in a phosphorus value of 2.55 mg/dl accompanied by a data flag indicating atypical kinetic behavior.The sample was then repeated on a second analyzer at decreased sample volume, resulting in a phosphorus value of 3.54 mg/dl.This value was deemed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer stated the patient sample was icteric upon visual check.The issue was resolved by re-drawing the patient.The investigation determined the customer actions resolved the issue.
|
| |
|