AMBU A/S ASCOPE 4 BRONCHO REGULAR; FLEXIBLE VIDEO BRONCHOSCOPE Back to Search Results
Model Number 477001000
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
An ascope 4 broncho regular bronchoscope was used instead of the slim.During use, the tip of the bronchoscope was teared inside the patient's tubus.The customer noted that this happened due to improper use (not according to the manufacturer's instructions).Patient outcome was not affected.
 
Additional Manufacturer Narrative
The defective device has not been returned for investigation.Neither sample nor lot number were available, therefore it has not been possible to retrieve a retention sample for investigation.Hence, it is not possible to verify the reported failure.Based on user feedback, the endoscope broke due to human error because ascope4 broncho regular was chosen by mistake instead of ascope4 broncho slim.Since there is no further information on the ett size used during procedure, it is suspected that the scope size was not compatible with the used et tube.Instructions for use (ifu) section 2.2 product compatibility informs which sizes of the ascope 4 broncho family is compatible with the different et tubes.Ambu will keep monitoring this issue and take actions if necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCOPE 4 BRONCHO REGULAR
Common Device Name
FLEXIBLE VIDEO BRONCHOSCOPE
Manufacturer (Section D)
AMBU A/S
baltorpbakken 13
ballerup, dnk 2750
DA  2750
MDR Report Key14985797
Report Number1220828-2022-00010
Device Sequence Number11162238
Product Code EOQ
UDI-Device Identifier05707480135509
UDI-Public05707480135509
Combination Product (Y/N)N
Initial Reporter CountryIS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number477001000
Device Catalogue Number477001000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/15/2022
Event Location Other
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/12/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-