• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MORTARA INSTRUMENT ELI380 LCD WAM WLAN USB DICOM AHA BAN; ELECTROCARDIOGRAPH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MORTARA INSTRUMENT ELI380 LCD WAM WLAN USB DICOM AHA BAN; ELECTROCARDIOGRAPH Back to Search Results
Model Number ELI380-ACX11
Medical Device Problem Code Intermittent Communication Failure (4038)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The eli 380 device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.If the eli 380 device loses connection to the wireless network, the device is unable to transmit or receive ekg¿s.Based on an analysis of the risk file, a failed connection on an eli380 may result in a delay of patient treatment which may cause or contribute to a serious injury or death.Therefore, hillrom is reporting this alleged connection failure as a product malfunction.The customer returned the device for investigation where it was determined that the issue was due to a corrupted pcba mainboard.The hillrom technician replaced the pcba motherboard of the eli380 device to correct the reported issue.Based on this information, no further action is required at this time.
 
Event or Problem Description
The customer reported that they were having trouble staying connected to the network using the eli380.There was no patient / user injury reported.This complaint was filed under hill-rom ref # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELI380 LCD WAM WLAN USB DICOM AHA BAN
Common Device Name
ELECTROCARDIOGRAPH
Manufacturer (Section D)
MORTARA INSTRUMENT
7865 n 86th street
milwaukee WI 53224
Manufacturer Contact
keighley crosthwaite
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14986628
Report Number2183461-2022-00032
Device Sequence Number1674523
Product Code DPS
UDI-Device Identifier00812345024531
UDI-Public812345024531
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberELI380-ACX11
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2022
Initial Report FDA Received Date07/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-