WRIGHTS LANE SYNTHES USA PRODUCTS LLC COMPRESSION/DISTRACTION INSTRUMENT; INSTRUMENT, COMPRESSION
|
Back to Search Results |
|
| Model Number |
03.111.021 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on an unknown date, the pin was bent inside the distractor, and it could not be removed.There was patient involvement.There were no patient outcome.This report is for one (1) compression/distraction instrument.This is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Occupation: reporter is a j&j employee.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the compression/distraction instrument had a fragment of an unknown device stuck in the insertion hole of the strut.Additionally, the body retainer sub-assembly is missing from the returned device.A dimensional inspection for the compression/distraction instrument was unable to be performed due to complex geometry of the device.A functional test was unable to be performed due to stuck components.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the compression/distraction instrument would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h6: a device history record (dhr) review was conducted: part number: 03.111.021.Lot number: t942754.Manufacturing site: tuttlingen.Release to warehouse date: 08-feb-2010.A review of the device history records was performed for the finished device lot number, and no non-conformances were identified.The raw material certificate was reviewed and the used material was according to the specification of the device.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|