| Model Number |
FD107 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A steris service technician, steris account manager and steris clinical educator arrived onsite following the reported event.User facility personnel were unable to identify the specific manifold rack that the employee was handling.The technician, account manager and clinical educator inspected all 20 manifold racks at the user facility and no rough edges or burrs were noted.The manifold racks were operating to specification and returned to service.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported that an employee obtained a cut to their finger while manually pulling a manifold rack out of their washer.The employee sought and received medical treatment (stitches).
|
| |
|