ST PAUL PNEUPAC PARAPAC MEDIC MODEL; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number 120002
Medical Device Problem Codes Defective Component (2292); Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventilator's knobs were broken and the device was not reaching pressure.No patient injury, adverse affects, or patient involvement were reported.
 
Additional Manufacturer Narrative
Other, other text: the device was returned for evaluation.The knobs were not broken but loose on the spindle.All pressure measurements were within specification, but the manometer was not stable due to the lack of maintenance by the user.During functional checks, additional issues were found; the demand valve was faulty, and the oxygen concentration was measuring too high.However, the primary problem was the demand body assembly, and replacement was made.Additional repairs were performed.Preventive maintenance was completed.No previous service history was found.The manufacturing device history record review is not relevant to the investigation due to the age of the device and the problem source being user interface.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC MEDIC MODEL
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14986979
Report Number3012307300-2022-13261
Device Sequence Number19266821
Product Code BTL
UDI-Device Identifier10610586045479
UDI-Public10610586045479
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K020899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number120002
Device Catalogue Number120002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer07/20/2022
11/08/2022
Initial Report FDA Received Date07/12/2022
Supplement Report FDA Received Date08/19/2022
11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/1980
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-