| Model Number |
2881201400 |
| Health Effect - Clinical Code |
Sepsis (2067)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the information available, the end-user was contacted on her 30-day follow up and she reported that, while on a 3-week trip to (b)(6) with her daughter she had been treated for urosepsis and spent 3 days in the hospital.The end-user started experiencing uti symptoms which was occasional pain, burning sensation and cloudy urine when she urinated.She went to a clinic in (b)(6) but was informed it was probably nothing.About 5-6 days after experiencing the symptoms she began to feel incredibly fatigued to the point of almost passing out.She was taken to the emergency room by her daughter where she was treated for urosepsis.The end-user will not use the catheters before seeing her doctor.The end-user had been properly sanitizing her vaginal area before and after catheterization.No malfunction to the catheter was reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
12 closed samples of this lot were tested.All samples met production requirements.The quality engineer checked the manufacturing process related to the given finished goods lot number and no nonconformities were found that were related to this event.A review of the production documents did not reveal any non-conformity that are related to this event.
|
| |
|